titles_ncdp


                              Titles:  NCDP
                              =============
 
Contact:  Wan, Oxford.
 
Revision History:-
================
 
5.00    H.S. Wan        First version.
        N. Oblath       Reorganization, use this bank for current pulse transformations. 
                        Add anode voltage
5.02    N. Oblath       Add Shaper, MUX and Scope dead/open times
5.03    N. Oblath       Add shaper smearing and direct-pulse fraction
        N. Oblath       Add transmission line parameters
	H.S. Wan        Add preamp reflection and noise scaling constants.
        N. Oblath       Add shaper noise (physics) constant; modify noise comments
5.03    N. Oblath       Add flag to check set thresholds


 
Titles Files
------------
 
ncd_input.dat
 
 
Verification
------------
 
 
Description
-----------
    
   This bank contains the quantities for electronics simulation.  
   (NCDP = NCD Properties bank) 
 
Data Words
----------
 
Line      Type   Variable   [symbolic command]   Description
----      ----   --------   --------------------------------

  1         I               Custom electronics parameters flag
  2         I               Check set thresholds flag
  3..15     I               Spare
  16        F               sqrt(LC) of NCD line (ns/cm, inverse of propagation velocity)
  17        F               * RC time constant of line for reflection simulation (ns)
  18        F               * Pre-amplifier gain (V/A)
  19        F               * Pre-amplifier low pass RC (ns)
  20        F               * Pre-amplifier high pass RC (ns)
  21        F               * MUX1 RC (ns)
  22        F               * MUX2 RC (ns)
  23        F               * Logamp constant a
  24        F               * Logamp constant b
  25        F               * Logamp constant c
  26        F               * Logamp time delay
  27        F               * Pretrigger offset
  28        F               Shaper threshold
  29        F               Shaper gain
  30        F               Mux threshold
  31        F               * Delays due to delay line (ns)
  32        F               Length of delay line (cm).
  33        F               * Cable RC (ns)
  34        F               * Counter RC (ns/cm)
  35        F               * Anode voltage 
  36        F               Shaper open time (ns)
  37        F               Shaper dead time (ns)
  38        F               MUX dead time (s; NOTE UNITS!)
  39        F               Scope dead time (s; NOTE UNITS!)
  40        F               Electronics sigma for shaper noise (full pulse sim.)
  41        F               Total sigma for shaper noise (shaper-only sim.)
  42        F               Direct-pulse fraction (1-reflected-pulse fraction)
  43        F               NCD Lossy Transmission Line resistance (Ohm/cm)
  44        F               NCD LTL inductance (H/cm)
  45        F               NCD LTL capacitance (F/cm)
  46        F               NCD LTL conductance (S/cm/MHz)
  47        F               Delay line LTL resistance (Ohm/cm)
  48        F               Delay line LTL inductance (H/cm)
  49        F               Delay line LTL capacitance (F/cm)
  50        F               Delay line LTL conductance (S/cm/MHz)
  51        F               Direct pulse fraction at preamplifier
  52        F               Noise scaling factor (mux bus0)
  53        F               Noise scaling factor (mux bus1)
  54        F               Noise scaling factor (mux bus2)
  55        F               Noise scaling factor (mux bus3)
  56        F               Physics sigma for shaper noise (full pulse sim.; ~adc^-0.5)

Notes
-----

* Quantities marked with an asterisk are nominally read from the 
NCLA and NCLB banks.  If desired, they can be read from this bank 
instead by using the "$custom_ncd_electronics $on" command in your 
command file, and setting the desired parameters to something 
other than -9999.  If the values are still -9999., they will be 
read from the NCLA/B bank.


Go Back to the Snoman Companion Top Page